The Wireless Analogue Output Modules provide a configurable analog output for a single T24 radio telemetry acquisition module. The output can be selected from the pre-calibrated voltage and current ranges of 0-10Volts, ±10Volts, 0-5Volts, ±5Volts, 0-20mA and 4-20mA (All current outputs both sink and source).
The LED indictors on the T24-AO1 provide status indication of the condition of the radio link, remote battery life and remote sensor status.
These Wireless Analogue Output Modules are configured and scaled using free T24 Toolkit software which runs on a PC connected to a T24 base station.
The T24-AO1 industrial version (T24-AO1i) is housed in an IP67 sealed enclosure and offers a digital input allows for zeroing of the incoming data value. This version also offers extended range over the desktop version.
The license free 2.4GHz direct sequence spread spectrum (DSSS) radio technology offers high integrity, error free communications which can co-exist with other wireless technologies such as Wi-Fi, Bluetooth® and Zigbee®. FCC, ETSI and IC approvals.
General Radio | ||||
---|---|---|---|---|
Parameter | Min | Typical | Max | Units |
Licence | Licence Exempt | |||
Modulation Method | MS (QPSK) | |||
Radio Type | Transceiver (2 way) | |||
Data Type | 250 | K bits/sec | ||
Radio Frequency | 2.4000 | 2.4835 | GHz | |
Power | 1 | mW | ||
T24-AO1 | 20 (65) | Metres (feet) * | ||
T24-AO1i | 800 (2600) | Metres (feet) * | ||
Channels (DSSS) | 16 |
* Maximum range achieved in open field site with the T24 module at a height of 3 metres (9.8 feet) above ground.
Parameter | Min | Typical | Max | Units | Notes |
---|---|---|---|---|---|
IP Rating : T24-AO1 | IP50 | ||||
IP Rating : T24-AO1i | IP67 | ||||
Humidity | 0 | - | 95 | %RH | |
External Supply Voltage Range | 9 | 12 | 36 | Volts | |
Operating Current | - | 85 | 150 | mA | |
Operating Temperature Range | -40 | - | 85 | Deg C | |
Storage Temperature Range | -40 | - | 85 | Deg C | |
Reverse Polarity Protection | - | - | -36 | Volts | Max supply level |
Digital Output Drive Voltage | 30 | Volts | |||
Digital Output Drive Current | 20 | mA | |||
Digital Input Load Impedance | 100 | Ohms | |||
Voltage Output | |||||
Resolution | - | 16 | Bits | ||
Output Gain Stability | - | 0.008 | 0.015 | ±%FS/°C | |
Output Zero Stability (1hr) | - | 0.006 | 0.015 | ±%FS/°C | |
Long Term Stability (10k hrs) | - | 0.003 | 0.01 | ±% FS | |
Residule Ripple | - | 40 | mV p-p | ||
Linearity | 500 | Ohms | |||
Bandwidth (-3dB) | TBD | Hz | |||
Current Output | |||||
Resolution | - | 16 | Bits | ||
4-20mA Output Gain Stability | - | 0.0060 | 0.0120 | ±%FS/°C | |
4-20mA Output Zero Stability | - | 0.005 | 0.0100 | ±%FS/°C | |
Short Term Stability (1hr) | - | 0.006 | 0.018 | ±%FS | |
Long term Stability (10k hrs) | - | 0.06 | 0.2 | ±%FS | |
Residule Ripple | 0.032 | mA p-p | |||
Settling Time to ±0.5uA (thermal effects) | - | 5 | - | secs | |
Maximum Load Impedance | 500 | Ohms | |||
Linearity | - | 0.01 | 0.02 | ±%FS | |
Bandwidth (-3dB) | TBA | Hz |
T24-AO1 | Code |
---|---|
Desktop Analogue Output Module | T24-AO1 |
Industrial Analogue Output Module | T24-AO1i |
File | Date | Size | |
---|---|---|---|
Wireless Analogue Output Module T24-AO1 datasheet (English) | 28/08/2015 | 656KB | |
Wireless Analogue Output Module T24-AO1 operating manual (English) | 28/08/2015 | 1022KB | |
T24 telemetry systems user's manual (English) | 26/03/2021 | 12.3MB | |
Application note AN200421A - Use of T24-ACM-PA, T24-AO1 and STU100-12V for remote speed monitoring (Portuguese) | 26/04/2021 | 982KB | |
T24 Toolkit v02.08.04 software - Configuration, calibration and testing of T24 range telemetry products | 10/09/2019 | 4.8MB | |
T24 Windows Drivers - Development files (DLLs, COM drivers - ActiveX, .NET, LabView) | 16/10/2018 | 6.2MB |