DSC is a high speed and high precision digitizing load cell conditioner, along with linearization and temperature compensation.
It was the ability to convert standard strain gauge load cells into digital load cells. Available output formats include the RS232, RS485, ASCII, Modbus, CAN and MantraCAN (J1939).
The digital load cell converter has the option of being housed in a convenient IP65 enclosure, or for multiple load cell installations, refer to models DSJ1 display & junction box for a single digital load cell for 1 channel, and DSJ4 for 4 channel.
DSCH High Stability | |||||
---|---|---|---|---|---|
Description | Min | Typical | Max | Units | |
Bridge Excitation | 4.5 | 5 | 5.25 | VDC | |
Sensor Impedance up to 18V Supply | 320 | 350 | 5,000 | Ohms** | |
Sensor Impedance up to 12V Supply | 120 | 350 | 5,000 | Ohms** | |
Bridge Sensitivity | -3 | +3 | mV/V | ||
Offset Temperature Stability | 1 | 4 | ppm/C | ||
Gain Temperature Stability | 3 | 5 | ppm/C | ||
Offset Stability with Time | 0.002 | 0.008 | %FR | ||
Gain Stability with Time | 30 | ppm of FR/1st Year | |||
Non Linearity | 0.0005 | 0.0025 | %FR | ||
Internal Resolution | 16 Million | Counts/Divisions | |||
Resolution @ 1Hz (Noise Stable) * | 200,000 | Counts/Divisions | |||
Resolution @ 10Hz (Noise Stable) * | 120,000 | Counts/Divisions | |||
Resolution @ 100Hz (Noise Stable) * | 50,000 | Counts/Divisions | |||
Resolution @ 500Hz (Noise Stable) * | 18,000 | Counts/Divisions | |||
Optional Temperature Measurement Resolution | 0.1 | ºC | |||
Optional Temperature Measurement Accuracy | 1 | ºC | |||
Notes: From original offset at any time. *Stability over 100 second period. **Subject to supply voltage. See Electrical Specification | |||||
Electrical | |||||
Power Supply Voltage | 5.6 | 12 | 18 | V dc | |
Power Supply Noise/Ripple | 100 | mV ac pk-pk | |||
Power Supply Current (350R Bridge) | 45 | 60 | mA | ||
Power @ 10V Supply (350R Bridge) | 350 | mW | |||
Sensor Impedance up to 18v Supply | 320 | 350 | 5,000 | ohms | |
Sensor Impedance up to 12v Supply | 120 | 350 | 5,000 | ohms | |
Excitation System | 4 wire | ||||
Environmental | |||||
Operating Temperature Range | -40 | 85 | °C | ||
Operating Temperature Range for OIML 6000d | -10 | 55 | °C | ||
Storage Temperature | -40 | 85 | °C | ||
Humidity | 0 | 95 | %RH on Condensing | ||
Communications | |||||
RS485 Data Rate | 2,400 | 230k | Baud | ||
CAN Bit Rate | 10k | 1M | Bits/Sec | ||
Note: Update speeds are selectable to 1, 2, 5, 10, 20, 50, 60, 100, 200, 300, 500 Samples/Sec | |||||
Protocols | |||||
Available Now: ASCII, MANTRABUS II, Modbus RTU, MantraCAN, Profibus - Contact Mantracourt for details. Please Note: MantraCAN products will only work with Mantracourt software up to and including Windows Vista. |
DSCS Industrial Stability | |||||
---|---|---|---|---|---|
Description | Min | Typical | Max | Units | |
Bridge Excitation | 4.5 | 5 | 5.25 | VDC | |
Sensor Impedance up to 18v Supply | 320 | 350 | 5,000 | Ohms** | |
Sensor Impedance up to 12v Supply | 120 | 350 | 5,000 | Ohms** | |
Bridge Sensitivity | -3 | +3 | mV/V | ||
Offset Temperature Stability | 5 | 10 | ppm/C | ||
Gain Temperature Stability | 30 | 50 | ppm/C | ||
Offset Stability with Time | 0.0035 | 0.016 | %FR „] | ||
Gain Stability with Time | 300 | ppm of FR /1st Year | |||
Non Linearity | 0.0005 | 0.0025 | % FR | ||
Internal Resolution | 16 Million | Counts/Divisions | |||
Resolution @ 1Hz (Noise Stable) * | 66,000 | Counts/Divisions | |||
Resolution @ 10Hz (Noise Stable) * | 40,000 | Counts/Divisions | |||
Resolution @ 100Hz (Noise Stable) * | 10,000 | Counts/Divisions | |||
Resolution @ 500Hz (Noise Stable) * | 5,000 | Counts/Divisions | |||
Optional Temperature Measurement Resolution | 0.1 | ºC | |||
Optional Temperature Measurement Accuracy | 1 | ºC | |||
Notes: From original offset at any time. *Stability over 100 second period. ** Subject to supply voltage. See Electrical Specification | |||||
Electrical | |||||
Power Supply Voltage | 5.6 | 12 | 18 | V dc | |
Power Supply Noise/Ripple | 100 | mV ac pk-pk | |||
Power Supply Current (350R Bridge) | 45 | 60 | mA | ||
Power @ 10V Supply (350R Bridge) | 350 | mW | |||
Sensor Impedance up to 18v Supply | 320 | 350 | 5,000 | ohms | |
Sensor Impedance up to 12v Supply | 120 | 350 | 5,000 | ohms | |
Excitation System | 4 wire | ||||
Environmental | |||||
Operating temperature range | -40 | 85 | °C | ||
Storage temperature | -40 | 85 | °C | ||
Humidity | 0 | 95 | %RH Non Condensing | ||
Communication | |||||
RS485 Data Rate | 2,400 | 230k | Baud | ||
CAN Bit Rate | 10k | 1M | Bits/Sec | ||
Note: Update speeds are selectable to 1, 2, 5, 10, 20, 50, 60, 100, 200, 300, 500 Samples/Sec | |||||
Protocols | |||||
Available Now: ASCII, MANTRABUS II, Modbus RTU, MantraCAN, Profibus - Contact Mantracourt for details. Please Note: MantraCAN products will only work with Mantracourt software up to and including Windows Vista. | |||||
Support Modules | |||||
Instrument Explorer Software free PC software, RS232 - RS485 Convertor, DSC Evaluation Kits | |||||
Environmental | |||||
CE Environmental Approvals | European EMC Directive 2004/108/EC, Low Voltage Directive 2006/95/EC |
High Stability | Code |
---|---|
RS232 output, ASCII Protocol | DSCH2ASC |
RS232 output, MANTRABUS Protocol | DSCH2MAN |
RS232 output, MODBUS Protocol | DSCH2MOD |
RS485 output, ASCII Protocol | DSCH4ASC |
RS485 output, MANTRABUS Protocol | DSCH4MAN |
RS485 output, MODBUS Protocol | DSCH4MOD |
MantraCAN Protocol | DSCHMCAN |
Industrial Stability | Code |
RS232 output, ASCII Protocol | DSCS2ASC |
RS232 output, MANTRABUS Protocol | DSCS2MAN |
RS232 output, MODBUS Protocol | DSCS2MOD |
RS485 output, ASCII Protocol | DSCS4ASC |
RS485 output, MANTRABUS Protocol | DSCS4MAN |
RS485 output, MODBUS Protocol | DSCS4MOD |
MantraCAN Protocol | DSCSMCAN |
Accessories | Code |
Digital temperature sensor module for wiring to DCell/DSC | DTEMP |
Evaluation Kit for RS232 | EVAL KIT 232 |
Evaluation Kit for RS485 | EVAL KIT 485 |
Evaluation Kit for CAN | EVAL KIT CAN |
Enclosures and options | Code |
1 Channel - DSJ1 PCB (PCB only) | DSJ1 |
1 Channel - IP65 ABS Grey Case | JAB |
4 Channels - DSJ4 PCB (PCB only) | DSJ4 |
Case and DIN Rail Mounting for DSJ4, LCA, UAB, JBA | D3 |
4 Channels - Transparent Plastic Case Lid | LTL |
4 Channels - IP65 ABS Grey Case | LAB |
4 Channels - Stainless Steel IP65 Case 220 x 160 x 90 mm | LSS |
4 Channels - Die Cast IP65 Case 220 x 120 x 80mm | LDC |